Product Description:
Conduct real-time central body temperature monitoring for patients during surgery.
Applicable for central body temperature monitoring in the esophagus, nasopharynx and rectum of adults;
Compatible with various monitors, providing accurate body temperature signals;
For single use to avoid the risk of cross-infection.
Product Performance:
1) Main performance: Used for collecting and transmitting physiological signals of patients' body temperature.
2) Applicable scope: Applicable to different models of multi-parameter monitors and continuous measurement electronic thermometers.
3) Temperature measurement range: 25 ºC - 45ºC Maximum allowable error: ±0.1 ºC
4) Response time: < 150s
5) Classification by degree of electric shock: According to the general classification rules for electric shock protection, the thermometer belongs to type BF, and the application part of the body temperature probe after matching also belongs to type BF.
6) Classification by degree of protection against liquid ingress: Ordinary equipment
7) Classification by operation mode: Continuous operation
8) Classification by safety degree when used in a flammable anesthetic gas mixed with air or a flammable anesthetic gas mixed with oxygen or nitrous oxide: Non-AP/APG.
9) Electromagnetic compatibility: Meets the requirements of YY 0505-2012.
10) Probe (resistance type) maximum dissipated energy: When immersed in a water tank at a temperature of 37 ± 0.1°C, the provided maximum power of 0.06mW should not cause energy consumption that leads to a temperature rise exceeding 0.02°C.
11) Stability: After placing the temperature probe in an environment of (80 ± 2) °C for 96h, the maximum allowable error value before and after probe exposure meets the requirements of ±0.1 ºC.
12) Body fluid protection: The insulation of the body temperature probe is strong enough to prevent the indicated temperature change from exceeding ±0.02°C when immersed in conductive liquid.
13) Resistance to cleaning and disinfection: After the body temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the manual, the maximum allowable error meets the requirements of ±0.1 ºC.