Location:
Disposable adult temperature measurement catheter semi-product (TE)

Disposable adult temperature measurement catheter semi-product (TE)

QR code
share
WeChat
Sina Weibo
QQ
QQ space
Douban
Baidu Post Bar
Disposable children's intracavitary body temperature sensor
Real-time central body temperature monitoring for intraoperative patients.
Suitable for central body temperature monitoring in children's esophagus, nasopharynx and rectum;
Compatible with various monitors, providing accurate body temperature signals;
Disposable to avoid the risk of cross-infection.
+86-755-26641550
Enquiry
Detail



Product description:
Real-time central body temperature monitoring for patients during surgery.
Suitable for central body temperature monitoring in children's esophagus, nasopharynx and rectum;
Compatible with multiple monitors, providing accurate body temperature signals;
Disposable to avoid the risk of cross-infection.


Product performance:

1) Main performance: For collecting and transmitting the physiological signals of patients' body temperature.

2) Applicable scope: Applicable to different models of multi-parameter monitors and continuous measurement electronic thermometers.

3) Temperature measurement range: 25 ºC - 45 ºC Max allowable error: ±0.1 ºC

4) Response time: < 150s

5) Classification by degree of electric shock: According to the general electric shock protection classification rules, the thermometer belongs to type BF. After the body temperature probe is matched with it, its application part also belongs to type BF.

6) Classification by the degree of protection against liquid ingress: Ordinary equipment

7) Classification by operation mode: Continuous operation

8) Classification by the degree of safety when used in the case of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: Non-AP/APG.

9) Electromagnetic compatibility: Meeting the requirements of YY 0505-2012.

10) Probe (resistance type) Maximum dissipated energy: When immersed in a water tank at a temperature of 37 ± 0.1 °C, the maximum power provided of 0.06mW should not cause energy consumption that causes the temperature to rise by more than 0.02 °C.

11) Stability: After placing the temperature probe in an environment of (80 ± 2) °C for 96 hours, the maximum allowable error value before and after the probe exposure meets the requirements of ±0.1 ºC.

12) Body fluid protection: The insulation of the body temperature probe is strong enough to prevent the indicated temperature from changing by more than ±0.02 °C when immersed in the conductive liquid.

13) Resistance to cleaning and disinfection: After the body temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the manual, the maximum allowable error meets the requirements of ±0.1 ºC.

Tel.:+86-0755-26641550
Email:sales01@jcr-med.com