Real-time body surface temperature monitoring for intraoperative patients.
Suitable for monitoring body skin and axillary temperature;
Compatible with multiple monitors, providing accurate temperature signals;
Reusable
+86-755-26641550
Enquiry
Detail

Product description:

Conduct real-time body surface temperature monitoring for patients during surgery.

Applicable for body skin and axillary temperature monitoring;

Compatible with multiple monitors, providing accurate body temperature signals;

Reusable


Product performance:

1) Main performance: Used for collecting and transmitting physiological signals of patients' body temperature.

2) Application scope: Applicable to different models of multi-parameter monitors and continuous measurement electronic thermometers.

3) Temperature measurement range: 32 ºC - 44 ºC Maximum allowable error: ±0.1 ºC

4) Response time: < 150s

5) Classification by degree of electric shock: According to the general classification rules of electric shock protection, the thermometer belongs to type BF, and the application part after the body temperature probe is matched also belongs to type BF.

6) Classification by degree of protection against liquid ingress: Ordinary equipment

7) Classification by operation mode: Continuous operation

8) Classification by safety degree when used in a situation with flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: Non-AP/APG.

9) Electromagnetic compatibility: Conforms to the requirements of YY 0505-2012.

10) Probe (resistance type) maximum dissipated energy: When immersed in a water tank with a temperature of 37 ± 0.1 °C, the provided maximum power of 0.06 mW should not lead to an energy consumption that causes the temperature to rise by more than 0.02 °C.

11) Stability: After placing the temperature probe in an environment of (80 ± 2) °C for 96 hours, the maximum allowable error value before and after the probe exposure conforms to the requirement of ± 0.1 ºC.

12) Body fluid protection: The insulation of the body temperature probe is strong enough to prevent the indicated temperature change from exceeding ± 0.02 °C when immersed in the conductive liquid.

13) Resistance to cleaning and disinfection: After the body temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the manual, the maximum allowable error conforms to the requirement of ± 0.1 ºC.