Product description:
Real-time central body temperature monitoring for intraoperative patients.
Suitable for central body temperature monitoring in the esophagus, nasopharynx and rectum of children;
Compatible with multiple monitors, providing accurate body temperature signals;
Disposable to avoid the risk of cross-infection.
Product performance:
1) Main performance: Used for collecting and transmitting the physiological signals of patients' body temperature.
2) Scope of application: Applicable to different models of multi-parameter monitors and continuous measurement electronic thermometers.
3) Temperature measurement range: 32 ºC - 44 ºC Maximum allowable error: ±0.1 ºC
4) Response time: < 150s
5) Classification by degree of electric shock: According to the general electric shock protection classification rules, the thermometer belongs to the BF type, and the application part of the body temperature probe after matching with it also belongs to the BF type.
6) Classification by the degree of protection against liquid ingress: Ordinary equipment
7) Classification by operation mode: Continuous operation
8) Classification by safety degree when used in the case of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: Non-AP/APG.
9) Electromagnetic compatibility: Meets the requirements of YY 0505-2012.
10) Probe (resistance type) maximum dissipated energy: When immersed in a water tank with a temperature of 37 ± 0.1 °C, the provided maximum power of 0.06 mW should not cause energy consumption that causes the temperature to rise by more than 0.02 °C.
11) Stability: After placing the temperature probe in an environment of (80 ± 2) °C for 96 hours, the maximum allowable error value before and after the probe exposure meets the requirement of ±0.1 ºC.
12) Body fluid protection: The insulation of the body temperature probe is strong enough to prevent the indicated temperature change from exceeding ±0.02 °C when immersed in the conductive liquid.
13) Resistance to cleaning and disinfection: After the body temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the manual, the maximum allowable error meets the requirement of ±0.1 ºC.