This product is used for one-time collection and transmission of temperature signals from the skin surface of the patient's oral cavity, rectum or armpit, and is matched with multi-parameter monitors or electronic clinical thermometers. This product is only for use by experienced medical professionals or under their guidance.
+86-755-26641550
Enquiry
Detail

Working principle, function and structure:

Place the sensor of the disposable temperature probe in the mouth or rectum or under the armpit and connect it to a compatible monitor or thermometer via a temperature extension cable. The main principle of this product is to use a stable resistance-temperature characteristic, a wide measurement range, and a resistance value that increases with temperature.

Reduced negative temperature coefficient thermistor.

There are 2 resistance values: 10K and 2.252K.


Product performance:

1) Main functions: to collect and transmit the patient's body temperature physiological signals.

2) Scope of application: Applicable to different models of multi-parameter monitors and electronic thermometers that can measure continuously.

3) Temperature measurement range: 25 ºC - 45 ºC Maximum allowable error: ±0.1 ºC

4) Response time: <150s

5) Classification by degree of electric shock: According to the general electric shock protection classification rules, the thermometer belongs to BF type, and the application part of the temperature probe also belongs to BF type when it is matched with it.

6) Classification by degree of protection against liquid ingress: General equipment

7) Classification by operation mode: Continuous operation

8) Classification by safety level when used with flammable anesthetic gas mixed with air or with flammable anesthetic gas mixed with oxygen or nitrous oxide: non-AP/APG.

9) Electromagnetic compatibility: Meet the requirements of YY 0505-2012.

10) Maximum dissipated energy of the probe (resistance type): When immersed in a water bath at a temperature of 37±0.1℃, the maximum power provided of 0.06mW should not result in energy consumption that causes the temperature to rise by more than 0.02℃.

11) Stability: After the temperature probe is placed in an environment of (80±2)℃ for 96 hours, the maximum allowable error value of the probe before and after exposure meets the requirement of ±0.1 ºC.

12) Body fluid protection: The insulation of the temperature probe is strong enough to prevent the indicated temperature from changing by more than ±0.02℃ when immersed in conductive fluid.

13) Resistance to cleaning and disinfection: After the temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the instructions, the maximum allowable error meets the requirement of ±0.1 ºC.