Product description:
Conduct real-time core body temperature monitoring for patients during surgery.
Suitable for core body temperature monitoring in children's esophagus, nasopharynx and rectum;
Compatible with multiple monitors, providing accurate body temperature signals;
Single-use to avoid the risk of cross-infection.
Product performance:
1) Main performance: Used for collecting and transmitting the physiological signals of patients' body temperature.
2) Scope of application: Applicable to different models of multi-parameter monitors and continuous measurement electronic thermometers.
3) Temperature measurement range: 32 ºC - 44 ºC Maximum allowable error: ±0.1 ºC
4) Response time: <150s
5) Classification by degree of electric shock: According to the general classification rules for electric shock protection, the thermometer belongs to type BF. After the body temperature probe is matched with it, its application part also belongs to type BF.
6) Classification by degree of protection against liquid ingress: Ordinary equipment
7) Classification by operation mode: Continuous operation
8) Classification by safety degree when used in the case of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: Non-AP/APG.
9) Electromagnetic compatibility: Complies with the requirements of YY 0505-2012.
10) Probe (resistance type) maximum dissipated energy: When immersed in a water tank with a temperature of 37 ± 0.1°C, the provided maximum power of 0.06mW should not cause energy consumption that results in a temperature increase exceeding 0.02°C.
11) Stability: After placing the temperature probe in an environment of (80 ± 2) °C for 96h, the maximum allowable error value before and after the probe exposure complies with the requirement of ±0.1 ºC.
12) Body fluid protection: The insulation of the body temperature probe is strong enough to prevent the indicated temperature from changing by more than ±0.02°C when immersed in the conductive liquid.
13) Resistance to cleaning and disinfection: After the body temperature probe is cleaned and disinfected according to the cleaning and disinfection methods recommended in the manual, the maximum allowable error complies with the requirement of ±0.1 ºC.